Calibration and repair of Dead Weight Tester

Calibration and repair of Dead Weight Tester

• Calibration should only be carried out when the pressure balance is in good working order.

• The operation of the pressure balance under calibration and the pressure reference standard should follow the laboratory’s calibration procedure and the manufacturer’s technical manual.

• Parameters should be controlled according to the uncertainty regime, including ambient temperature within 15 °C and 25 °C, and control of door openings and operator movement.

• Device installation should be done away from air disturbances, near the standard instrument, and using a rigid, stable table.

• The height difference between reference levels of the two instruments to be compared should be minimized.

• Respect the verticality of the piston as recommended by the manufacturer.

• Short, wide-bore pipework should be used, and the cleanliness and tightness of the tubing should be ensured.

• An appropriate drainage system should be installed to control the nature of the fluid in the tubing.

• A suitable temperature measurement system should be attached.

• For gas gauge pressure, use a clean and dry gas at a temperature near ambient.

• For gas absolute pressure, use a clean pump or an appropriate trap.

• For liquid pressure, use the liquid recommended by the manufacturer. If the liquid in the balance under calibration is different from the standard, use an appropriate interface/separator.

Calibration of Pressure Balances

• The standard pressure reference instrument for calibration is a pressure balance, with mercury column manometers used for ranges lower than 300 kPa.

• Calibration of an absolute pressure balance can be carried out in gauge mode with an added uncertainty in A0.

• Calibration of a pressure balance intended for measuring negative gauge pressures can be carried out in positive gauge pressure mode with an added uncertainty in A0.

• The reference instrument must be traceable to a national standard with a recognized calibration certificate and have an uncertainty better than the presupposed uncertainty of the balance to be calibrated.

 

 

 

 

Preparation of the Pressure Balance

• The pressure balance under calibration should be placed in the laboratory at least 12 hours before the calibration is started to reach thermal equilibrium.

• The piston-cylinder assembly should be cleaned and inspected for surface scratches and corrosion.

• The piston-cylinder assembly should be examined for free rotation time, descent rate, and connection to the standard instrument.

• The reference level for both pressure balances should be identified and reduced as much as possible.

• For absolute pressure, the difference in height between the reference level of the standard and the reference level of the balance to be calibrated should be measured.

• The piston or cylinder should be rotated according to the manufacturer’s recommendation.

Example of Calibration Procedure

• Both methods are comparative, comparing the balance to be calibrated and the standard instrument when both are submitted to the same pressure and environmental conditions.

Method B - Effective Area Determination Method

• The method determines the mass of all weights, including the piston of the pressure balance, if removable.

• The effective area of the piston-cylinder assembly of the pressure balance is determined as a function of pressure.

• The repeatability of the measured pressure is estimated from the experimental standard deviation calculated from successive determinations.

• The relative uncertainty of mass determination should not exceed 20% of the likely total measurement uncertainty of the pressure balance to be calibrated.

• For pressure balances equipped with both low and high pressure piston-cylinder assemblies or with removable piston-cylinder assemblies, the complete calibration process should be carried out for each piston-cylinder assembly.

• The repeatability of the measured pressure is estimated from the experimental standard deviation calculated from the successive determinations operated for each pressure point.

• The cross-floating procedure is carried out at each measuring point when using a pressure balance as a standard instrument.

Calibration of Pressure Balances

• Adjust the pressure to equilibrate the balance under calibration.

• Perform an adjustment with small weights on one of both instruments (usually the more sensitive to a change in mass) if method B is used, or only on the reference pressure balance if method A is used.

• The equilibrium condition should be reached when the proper falling rate of both pistons is found.

• Note the reference number of each of the weights applied on both balances.

• Note the temperature of the piston-cylinder assembly of both balances. If the balance is not equipped with a temperature probe, note the surrounding air temperature using an electronic thermometer attached to some suitable point of the balance.

• Alternatively, a differential pressure gauge can be used which is installed in the pressure line connecting two pressure balances under comparison.

• A high accuracy pressure gauge can be used if the measurement range of the gauge should cover the calibration pressure range.

• For each pressure point, adjust the pressure and masses to equilibrate the pressure balances.

• If the differential pressure is so high that the needed uncertainty cannot be reached from the calibration of the transducer, repeat the last five operations.

Absolute Dead weight tester Calibration

• Open bell jars and both valves.

• Place weights on both balances.

• Adjust pressure and masses to equilibrate balances.

• Close and evacuate bell jars.

• Adjust pressure on both sides to equilibrate balances.

• Record pressure gauge reading from reference balance.

• Repeat operations if difference is high.

• Note reference number of weights applied on both balances.

• Note piston-cylinder assembly temperature and surrounding air temperature if balance isn't equipped with a temperature probe.

• Note residual pressure in bell jar of both balances.

Uncertainty of measurement for dead weight tester

Method A:

• Estimates type A uncertainty (repeatability of the balance) based on the standard deviation of the pressure difference.

• Type B uncertainty (reference pressure, temperature, head correction, tilt, spin rate/direction, residual pressure, and limited cross-floating sensitivity):

• Standard uncertainty for each component is calculated in accordance with the JCGM 100:2008 (GUM).

 

Method B:

• Estimates type A uncertainty (repeatability of the balance) based on the standard deviation of the effective area.

• Estimates type B uncertainty (reference pressure, masses, temperature, thermal expansion coefficient, air buoyancy, head correction, surface tension of the pressure-transmitting fluid, tilt, spin rate/direction, residual pressure):

• Standard uncertainty for each component is calculated in compliance with the JCGM 100:2008 (GUM).